Catalog Number 0375951100
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device broke during procedure.No pieces fell into the joint space and the procedure was completed successfully.
 
Additional Manufacturer Narrative
Alleged failure: the rotating tube in the outer tube of the barrel burs is broken off during work.The failure identified in the investigation is consistent with the complaint record.The probable root cause/s could be the user applied an excessive force, continued use of pressure such as bending or prying may cause to shaft to bend and break.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the device broke during procedure.No pieces fell into the joint space and the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., BARREL BUR, 6 FLUTE F-SERIES - 5.0MM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8126970
Report Number0002936485-2018-00710
Device Sequence Number13761567
Product Code GFA
UDI-Device Identifier37613327054850
UDI-Public37613327054850
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0375951100
Device Lot Number18045CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date12/03/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-