Catalog Number 03.010.405
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2018, there was a functional issue with aiming arm, hole for locking bolt at proximal femoral nail antirotation (pfna) could not hit properly.There was no patient harm.There was no surgical delay reported.This report is for one (1) radiolucent insertion handle.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A device history record (dhr) review was conducted: part: 03.010.405 lot: l267342 manufacturing site: hägendorf release to warehouse date: 04.April 2017 the device history record shows this lot of 12 pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.A product investigation was conducted.Visual inspection: the returned parts aim-arm (03.010.407) and insert-handle (03.010.405) were together assembled.All in all, the parts are shown in used condition.The function test has shown that the proximal hole of femoral nail antirotation (pfna) fit properly as intended.No abnormal findings were identified on these returned parts.Function test: no abnormal findings were identified on these returned parts.The function test has shown that the proximal hole of femoral nail antirotation (pfna) fit properly as intended.Summary: the complaint condition is unconfirmed.The returned parts aim-arm (03.010.407) and insert-handle (03.010.405) were together assembled.All in all, the parts are shown in used condition.The function test has shown that the proximal hole of femoral nail antirotation (pfna) fit properly as intended.No abnormal findings were identified on these returned parts.Based on that and the condition of the parts a product related issue can be excluded.Unfortunately the exact cause which has led to the occurrence could not be evaluated.We only can assume that a mechanical overloading situation, for example strongly torsion (aim-arm against pfna nail) during implantation, or the nail was not tightened enough on the insert handle.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed as no product related issue could be detected device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIOLUCENT INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8127001
Report Number8030965-2018-58716
Device Sequence Number2165410
Product Code LXH
UDI-Device Identifier07611819349421
UDI-Public(01)07611819349421
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.405
Device Lot NumberL267342
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 11/14/2018
Supplement Date Received by Manufacturer12/18/2018
01/17/2019
Initial Report FDA Received Date12/03/2018
Supplement Report FDA Received Date01/11/2019
01/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-