| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pt info: unk.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -8.5 diopter, into the patient's left eye (os) on (b)(6) 2018.On (b)(6) 2018 the lens was exchanged with a shorter lens due to excessive vault.The problem is resolved."no patient injury" is reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned dry in a contact lens case.There was clear surgical residue on the lens surface.Visual inspection found the haptic deformed.(b)(4).
|
| |
|