Insulin pump received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime and compromised forced sensor system and excessive no delivery test or verify signal to low alarm and unexpected restart alarm no delivery alarm and motor error alarm due to blank display.The motor was tested outside of the device and passed.
|