MEDTRONIC MEXICO KYPHON XPANDER INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K09A
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: the reported complaint of a balloon rupture was confirmed during use.Visual examination identified tears near the proximal peak of the balloon.This observation is consistent with rupture due to contact with bone splinters during instrument usage.The above observations are consistent with intraoperative instrument damage.Additional balloon expander was sent back with an inflation issue.The balloon would not expand.It appeared to be clogged up.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, intra-op, the balloon burst at 200 psi.The procedure was completed with another device.There was a delay in the surgery.The device came in contact with the patient.No patient complications were occurred due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC MEXICO
avenida paseo cucapah # 10510
parque industrial ei lago
tijuana
MX 
MDR Report Key8128011
Report Number9612164-2018-03458
Device Sequence Number1244348
Product Code HRX
UDI-Device Identifier00643169103061
UDI-Public00643169103061
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK09A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Initial Report FDA Received Date12/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-