| Catalog Number |
123514A |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water or for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.".
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter leaked.The catheter was inflated with 10ml but 2 ml was leaking on the side.It was also difficult to deflate the catheter.
|
| |
|