MALLINCKRODT MEDICAL SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 60XLTIN
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code Extubate (2402)
Date of Event 10/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the patient came back from endoscopy and after the device was placed, the patient developed respiratory distress.A plug was suspected and they tried to bag and suction the patient.They were unable to pass the suction catheter.The existing inner cannula was removed and found to have a kink/deformation in the tube that severely limited the size of the inner lumen.Once a new inner cannula was placed they were able to bag and suction properly and return the patient to baseline.The original inner cannula was determined to be the cause due to deformation/kink.Nothing actually occurred with the tube.Only the inner cannula was noted as the cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key8128633
Report Number8020889-2018-00135
Device Sequence Number19224694
Product Code JOH
UDI-Device Identifier40884522006720
UDI-Public40884522006720
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K051416
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60XLTIN
Device Catalogue Number60XLTIN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Initial Report FDA Received Date12/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-