As reported, during implantation, a certas plus valve was observed leaking from the reservoir.It was replaced with a new valve, the procedure completed.The surgeon commented that the valve should not have been damaged during the procedure.No further information was provided by the hospital.The valve will be returned.
|
The device was returned for evaluation.The position of the cam when the valve was received was at setting 5.The valve was visually inspected; the silicone housing was cut/torn in the needle chamber and the needle guard was raised.The root cause for the damage silicone house is probably due to a sharp or pointed object coming into contact with the silicone.As noted in the ifu silicone has a low cut/tear resistance.The root cause for the raised needle guard base, was probably due to user error, as noted in the ifu: do not fold or bend the valve during insertion.Folding or bending might cause rupture of the silicone housing, needle guard dislodgement, or occlusion of the fluid pathway.A review of manufacturing records found that the device conformed to specification when released to stock.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|