• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number H965458830
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It has been indicated that the used catheter will be returned for evaluation, however it has not yet arrived.Upon receipt of the sample, and completion of the investigation, a supplemental medwatch will be submitted.((b)(4)).
 
Event or Problem Description
As reported by angiodynamics' distributor in (b)(4), indwelling bioflo picc was removed due to a cracked hub.The used catheter will be returned to angiodynamics for evaluation.No patient injury was reported.
 
Additional Manufacturer Narrative
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the bioflo picc product family and the failure mode "hub broken/cracked {pasv & hybrid}." no adverse trend was indicated.As received, the catheter was cut at the 32cm mark and contained unknown fluid inside and out.The hubs had a needleless connector (nc), not supplied in the reported kit, attached to the hubs.The female luer lock component of the purple valve was confirmed to be cracked just adjacent to the molded parting line.There were no other visual defects noted to the valve or catheter.An accurate verification/measurement could not be performed for the female taper and threads of the valve due to the crack.The likely root cause for the cracked hub is an over-tightened connection with a mating male luer (likely the needleless connector).Inadequate flushing of the device may also be a contributing factor.The directions for use (dfu) supplied with the reported picc device contains caution that "repeated overtightening may reduce hub connector life," and not to use hemostats to "secure or remove devices with luer lock hub connections." recommended flushing techniques are also stated in the dfu.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / BIOFLO
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key8129015
Report Number1317056-2018-00221
Device Sequence Number1770913
Product Code LJS
UDI-Device IdentifierH965458830
UDI-PublicH965458830
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121089
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Catalogue NumberH965458830
Device Lot Number5296439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2019
Initial Date Received by Manufacturer 11/06/2018
Supplement Date Received by Manufacturer11/06/2018
Initial Report FDA Received Date12/04/2018
Supplement Report FDA Received Date03/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-