Catalog Number 72200730
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that it takes a lot of effort for the bonecutter to take some of the bone.The bonecutter blade produced small chips of metal in the joint.They were removed by flushing out.A backup device was available to complete the procedure with no significant delay or patient injuries.
 
Additional Manufacturer Narrative
Date of the event is unknown.
 
Additional Manufacturer Narrative
Additional information in: expiration date: (b)(4) 2023 device manufacture date: (b)(4) 2018 correction.Catalog number: 72200730 unique identifier (udi) #: (b)(4).
 
Additional Manufacturer Narrative
One 72200730 dyonics 4.0mm acromioblaster burr product reported on.The complaint states: ¿it was reported that it takes a lot of effort for the bone cutter to take some of the bone.The bone cutter blade produced small chips of metal in the joint.They were removed by flushing out.A backup device was available to complete the procedure with no significant delay or patient injuries.¿ the product was not returned for evaluation.Due to product unavailability, the complaint could not be ultimately confirmed.Definitive conclusions, accurate investigation and evaluation were not fully possible without evaluation of physical product.Factors that may affect device performance include: device ability, surgical ability, implant location and tissue condition.Shedding material condition has been found to be consistent with the device having inadvertent pressure applied while deburring which causes wear bands on the inner blade surface.The symptom points to excessive lateral load during use causing a mis-alignment of the inner and outer blade resulting in the shedding.Excessive ¿side-loading¿ on the blade during use does not improve cutting performance and in extreme cases may result in wear and degradation of the inner assembly.Final product met predetermined specifications upon release to distribution.No patient harm or injury is being reported as a result of this device related incident, and all of the metal debris was flushed out.There is no indication of adverse patient consequence as a result.No further clinical assessment is warranted based on the information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 4.0MM ELITE ACROMIONIZER BURR(6)
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8129019
Report Number1219602-2018-01639
Device Sequence Number11152181
Product Code HAB
UDI-Device Identifier03596010578020
UDI-Public03596010578020
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2023
Device Catalogue Number72200730
Device Lot Number50718389
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/19/2018
Supplement Date Received by Manufacturer12/05/2018
03/18/2019
03/19/2019
Initial Report FDA Received Date12/04/2018
Supplement Report FDA Received Date12/10/2018
03/18/2019
03/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-