• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. HIGHLY CROSSED LINKED KNEE; EMPOWR PS KNEETM TIBIAL INSERT, SIZE 8, 19MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. HIGHLY CROSSED LINKED KNEE; EMPOWR PS KNEETM TIBIAL INSERT, SIZE 8, 19MM Back to Search Results
Model Number 130-03-738
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/06/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to instability and quadriceps repair.The surgeon removed and exchanged out the insert.Towards the end, the insert was not big enough, so the new insert was removed along with all original djo implants and replaced with competitor brand implants.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to instability.The previous surgery and the surgery detailed in this investigation occurred 2.2 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGHLY CROSSED LINKED KNEE
Common Device Name
EMPOWR PS KNEETM TIBIAL INSERT, SIZE 8, 19MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8129205
Report Number1644408-2018-01059
Device Sequence Number13520002
Product Code JWH
UDI-Device Identifier00888912108560
UDI-Public(01)00888912108560
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number130-03-738
Device Catalogue Number343-19-708
Device Lot Number056U1001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/06/2018
Supplement Date Received by Manufacturer01/11/2019
Initial Report FDA Received Date12/04/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
242-01-107, LOT 860T1003; 343-19-708, LOT 056U1001; 351-01-108, LOT 948R1041; 242-01-107, LOT 860T1003; 343-19-708, LOT 056U1001; 351-01-108, LOT 948R1041
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-