• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK VASCULAR INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK VASCULAR INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number G35536
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the advance 35 lp low profile balloon catheter was found to be kinked in the mid-part of the device while still in the package.Sharp edges were observed.Another brand of balloon catheter was used to successfully complete the angioplasty with stenting procedure.The complaint device never made patient contact.No adverse effects were experienced.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation evaluation.A review of the complaint history, device history record, instructions for use (ifu), and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, a review of the design history file and quality control procedures was conducted, and no gaps were discovered.Moreover, an ifu is provided with the device, which states instructions to inspect the device upon removing it from the package to ensure no damage as occurred.Based on the information provided, no product returned and the results of our investigation, investigation has concluded that this device was likely damaged during transport or storage of the device.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK VASCULAR INC
1186 montgomery lane
vandergrift PA 15690
MDR Report Key8129286
Report Number1820334-2018-03503
Device Sequence Number1569166
Product Code LIT
UDI-Device Identifier10827002355368
UDI-Public(01)10827002355368(17)201208(10)NS8429903
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K132020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date12/08/2020
Device Model NumberG35536
Device Catalogue NumberPTA5-35-135-6-20.0
Device Lot NumberNS8429903
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/26/2018
Supplement Date Received by Manufacturer02/08/2019
Initial Report FDA Received Date12/04/2018
Supplement Report FDA Received Date02/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-