| Catalog Number |
0375941012 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.Please note all pieces were retrieved and the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: top portion of the burr broke into the patient's shoulder the failure identified in the investigation is consistent with the complaint record.The probable root cause/s could be an inadequate welding process, a bad weld location, probable cause could be part length variation, set-up and operator error.The blade comes in contact with a hard object, excessive pressure.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.Please note all pieces were retrieved and the procedure was completed successfully.
|
| |
|