Product complaint # (b)(4).Date sent to the fda: 01/02/2019.Investigation summary: it was reported suture breakage.An empty opened foil, a needle and three suture pieces of product code d7190, lot # mgz197 were returned for analysis.During the visual inspection of sample, the swage and attachment area were noted to be as expected and a suture piece was noted attach to barrel hole, the suture has begun with the process of degradation and the strand was broken in multiples pieces, this is the cause that breakage suture.The product code d7190 contains an absorbable suture and as the sample was received open, the time of exposure of suture to the environment could not be determined and no functional test can be performed due to sample condition.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition the assignable cause of the suture breakage was suture degradation exposure to environment.
|