| Catalog Number |
0006412 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Immucor technical support use a remote electronic connection method on (b)(6) 2018 to assess the instrument test well images in question, which were visually positive with red cell agglutinates.The camera images were optimal and aligned.The event log showed no errors during processing.The internal immucor number for this report is (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a customer site reported an unexpectedly positive outcome with anti-d (series 4) monoclonal blend by echo lumena instrument.
|
| |
|