• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANNULATED 2.5MM HEXAGONAL SCREWDRIVER SHAFT; SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANNULATED 2.5MM HEXAGONAL SCREWDRIVER SHAFT; SCREWDRIVERS Back to Search Results
Catalog Number 314.10
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter is synthes sales representative.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, after going through the decontamination process, there were several instruments inspected that appeared to be damaged.For the bone reduction forceps, the hinge is overly tight and very hard to open.The cannulated screwdriver tip, and cannulated power shaft tip both appear to be stripped.The 2.5mm hex power shafts tips also both look stripped.Eight (8) 2.8mm locking towers were inspected by the sales consultant which appeared to have damaged threads.The t4 star drive for the 1.5mm headless compression screw set was damaged and accidentally placed into the sharps container where the sales consultant could not retrieve it.There was no patient involvement.This report is for one (1) cannulated 2.5mm hexagonal screwdriver shaft.This is report 2 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: device evaluated by mfr, device manufacture date.Device history lot part: 314.100, lot: 8015166, manufacturing site: hägendorf, release to warehouse date: 04 sep 2012.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process review of the device history record(s) showed that there were no issues during the manufacture of this product, which would contribute to this complaint condition.Investigation summary: visual inspection: visual inspection on the returned screw driver observed worn at distal tip hex edges.No further issues were identified that would impact the function of the device.Conclusion: with the given age of the device and after a visual inspection, it is determined that the reusable instrument is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED 2.5MM HEXAGONAL SCREWDRIVER SHAFT
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8130600
Report Number2939274-2018-55269
Device Sequence Number16206190
Product Code HXX
UDI-Device Identifier10886982188502
UDI-Public(01)10886982188502
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.10
Device Lot Number8015166
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2018
Initial Date Received by Manufacturer 11/16/2018
Supplement Date Received by Manufacturer01/07/2019
01/14/2019
Initial Report FDA Received Date12/04/2018
Supplement Report FDA Received Date01/08/2019
01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-