BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00516940
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is received, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex esophageal partially covered stent was to be used to treat a 6 cm malignant stricture due esophageal cancer during an esophagogastroduodenoscopy with stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was tight and was not dilated prior to stent placement.According to the complainant, during the procedure, the stent was deployed in the esophagus.However, under fluoroscopy, it was noticed that the stent slipped into the stomach lumen.The stent was removed with a snare and a different device was used to complete the procedure.The patient was reported to be in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ESOPHAGEAL
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8130697
Report Number3005099803-2018-61682
Device Sequence Number13845343
Product Code ESW
UDI-Device Identifier08714729765288
UDI-Public08714729765288
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K073266
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2020
Device Model NumberM00516940
Device Catalogue Number1694
Device Lot Number0021915067
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date12/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-