Model Number EEG-1200A
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the eeg data recorded under another patient's study node.No patient harm or injury reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
The customer reported that the eeg data recorded under another patient's study node.No patient harm or injury reported.
 
Additional Manufacturer Narrative
H10: additional narrative: the customer reported that their system is recording under incorrect name.Service requested: troubleshooting.Service provided: troubleshooting.Investigation result: the last two studies started on customer's unit were saved under one name.The issue was resolved by ts, starting a new data schedule with the correct patient information, noted the db key and changed the nodes that were incorrectly filed under the wrong patient information.The root cause of the issue was unable to be determined.A review of device history found no previously reported issue with this unit.The customer's issue was resolved and no similar issue was reported subsequent to the resolution.The root cause of the issue could not be determined.Qe investigation for the complaint record has been completed.Corrected information: g4.Date received by manufacturer: should be 11/06/2018 not 12/04/2018 as listed on mdr initial report.Additional information: b4.Date of this report; e1.Initial reporter email; f6.Date user facility/importer became aware of the event; f7.Type of report; f11.Date report sent to fda; f13.Date report sent to manufacturer; g4.Date received by manufacturer; g7.Type of report; h2.If follow-up, what type? additional information correction; h3.Device evaluated by manufacturer? ; h6.Event problem and evaluation codes; h10.Additional manufacturer narrative.The following fields do not contain information as this unit is not serialized.Therefore, the following fields do not contain information.D4.Serial #.F9.Approximate age of device.H4.Device manufacture date.
 
Event or Problem Description
The customer reported that the eeg data recorded under another patient's study node.No patient harm or injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEG-1200A
Common Device Name
NEUROFAX ELECTROENCEPHALOGRAPH
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
MDR Report Key8131035
Report Number8030229-2018-00461
Device Sequence Number1579796
Product Code OLT
UDI-Device Identifier04931921110904
UDI-Public4931921110904
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K080546
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/01/2019
Report Date (Section F) 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEEG-1200A
Device Catalogue NumberEEG-1200A
Device Lot NumberNOT APPLICABLE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/31/2019
Event Location Hospital
Date Report to Manufacturer08/01/2019
Initial Date Received by Manufacturer 12/04/2018
Supplement Date Received by Manufacturer07/31/2019
Initial Report FDA Received Date12/04/2018
Supplement Report FDA Received Date08/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NOT APPLICABLE
Patient Sequence Number1
Date Report Sent to FDA08/01/2019
-
-