MEDTRONIC MINIMED PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-722LNAH
Medical Device Problem Codes No Display/Image (1183); Adverse Event Without Identified Device or Use Problem (2993); Power Problem (3010); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer's mother reported via phone call that the insulin pump was died and had a blank display.The customer¿s blood glucose level was unknown at the time of the incident.The customer was assisted with troubleshooting and declined for blank display.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan per healthcare professional instructions.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test at 0.08730 inches.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device powered up properly after the test battery was installed.Device was monitored for 4 days.No blank display anomaly or unexpected battery power loss anomaly noted.Device uploaded properly using carelink.No damage noted on the lcd glass.Device had minor scratched display window, cracked display window, scratched case, cracked case at display window corner, cracked battery tube threads, missing end cap sticker and a partially broken off reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8132132
Report Number2032227-2018-70908
Device Sequence Number10805047
Product Code OYC
UDI-Device Identifier00613994597236
UDI-Public(01)00613994597236
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-722LNAH
Device Catalogue NumberMMT-722LNAH
Device Lot NumberA000139231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2020
Initial Date Received by Manufacturer 11/27/2018
Supplement Date Received by Manufacturer05/05/2020
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date05/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age23 YR
Patient Weight73
-
-