• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY BCS II FEMORAL OXIN LT SZ 4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY BCS II FEMORAL OXIN LT SZ 4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Catalog Number 74022124
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Muscular Rigidity (1968)
Date of Event 05/20/2015
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient (b)(6) presented left knee ankylosis related to journey ii bcs retro implants.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, no clinical supporting documents were provided to conduct a thorough assessment of the reported issue.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches/failure mode.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Event or Problem Description
It was reported that patient (b)(6) presented left knee ankylosis related to journey ii bcs retro implants.Event was treated with a manipulation under anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY BCS II FEMORAL OXIN LT SZ 4
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8133010
Report Number1020279-2018-02681
Device Sequence Number1672614
Product Code JWH
UDI-Device Identifier00885556170366
UDI-Public00885556170366
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/11/2024
Device Catalogue Number74022124
Device Lot Number14LM09894
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/13/2018
Supplement Date Received by Manufacturer11/13/2018
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date02/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age67 YR
-
-