Model Number S-LLF200TG
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) confirmed that the devices met all material, assembly and performance specifications.Analysis of the device revealed that the total length of the fiber is 12 feet 0.5 inches, connector not included in over-all length.The connector is detached and returned.The fiber has 0.5 inches of the fiber optic exposed at the proximal end, which was original inside the connector.The fiber proximal end exhibits severe burn marks and melting.The distal tip of the fiber exhibits is fractured.Black discoloration was noted near the proximal end within blue jacket.The ferrule is broken within the connector and the core fiber is loose.The fiber optic is broken within the connector.An evaluation conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
Analysis of the return device revealed that the distal tip of the fiber is fractured.There were no patient complications reported due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFLEX FIBERS
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS INC.
3070 orchard drive
san jose CA 95134
MDR Report Key8134251
Report Number2937094-2018-60150
Device Sequence Number13761560
Product Code GEX
UDI-Device Identifier00878953004822
UDI-Public00878953004822
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K140679
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/03/2021
Device Model NumberS-LLF200TG
Device Catalogue NumberS-LLF200TG
Device Lot Number21917858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2018
Initial Report FDA Received Date12/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-