• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC SOFT-VU SIZING ANGIOGRAPHIC CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC SOFT-VU SIZING ANGIOGRAPHIC CATHETER Back to Search Results
Catalog Number H787107322055
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.Complaint # (b)(4).
 
Event or Problem Description
When prepping for an angiographic procedure and opening the sterile packaging of a softvu angiographic catheter, it was noted the tip of the catheter had fractured off inside of the package.The device was set aside and a new of the same was used to complete the procedure.There was no adverse patient effect due to this event.The reported defective disposable device has been returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was a soft vu catheter and the fractured off tip.A visual review of the device noted that the tip material exibits signs of cracking.The reported complaint description of tip fracture is confirmed.The root cause of this event was determined to be the material of the catheter tip.A material change to address the robustness of the catheter tip was implemented on (b)(6) 2017.The complaint sample sub-assembly was manufactured before the implementation of this material change.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with states; "never advance or retract an angiographic catheter or guidewire against resistance.This may result in damage to the vessel, the product, or both.Do not attempt to hand straighten the tip of any curved tipped catheters which are furnished with a tip straightener.This may result in damage to the product.Tip straighteners are furnished on all catheters intended to be straightened with the aid of a tip straightener.Always use a guidewire to remove the catheter from the vasculature.Failure to do so may result in damage to the vessel, puncture site, product, or all three.The maximum pressure limit of catheters intended for flush angiography is stated on the catheter package.When using a pressure injector, do not exceed the stated maximum pressure.Catheters intended for selective angiography do not have a maximum pressure limit stated on the catheter package.Typical flow rates of up to 10cc per second are stated with pressure generated to achieve these typical flow rates.Never advance or retract an angiographic catheter or guidewire against resistance.This may result in damage to the vessel, the product, or both.Angiodynamics angiographic catheters are designed for use with specific guidewire diameters.The recommended maximum guidewire diameter is specified on the catheter label".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFT-VU SIZING ANGIOGRAPHIC CATHETER
Common Device Name
SOFT-VU SIZING ANGIOGRAPHIC CATHETER
Manufacturer (Section D)
ANGIODYNAMICS, INC
603 queensbury avenue
queensbury NY 12804
MDR Report Key8134936
Report Number1319211-2018-00118
Device Sequence Number13874552
Product Code DQO
UDI-Device IdentifierH787107322055
UDI-PublicH787107322055
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K112452
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Catalogue NumberH787107322055
Device Lot Number5150779
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Supplement Date Received by Manufacturer11/12/2018
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date03/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-