| Model Number |
6947M72 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that a carealert indicated that the patient received inappropriate therapy after oversensing noise.There were other episodes that indicated one to two seconds of oversensing.It was also noted that approximately a week earlier the right ventricular (rv) lead had a lead integrity warning triggered due to high rate non-sustained episodes and sensing integrity counter (sic).The rv lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the noise on the right ventricular (rv) lead was cautery noise from a procedure done that day.
|
| |
|