| Catalog Number |
0006428 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The full number for the product in question is bk020053 (1/27/03).Immucor technical support used a remote electronic connection method on (b)(4) 2018 to assess the testing instrument in question.The test well images in question were visually negative.Because the product was already expired, the immucor laboratory subsequently performed a dhr review, and determined that the product had met all necessary specifications at the time of the product release.(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a customer site reported an unexpectedly negative antibody identification when using capture-r ready indicator red cells on a galileo echo instrument, when tested on (b)(6) 2018.
|
| |
|