• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(bathroom).Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with tests results from all the results obtained.The expected fasting blood glucose test result range is 90-100mg/dl.The customer did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bathroom.During the call on (b)(6)2018, a blood test was not performed by the customer.The test strip lot manufacturer's expiration date is 08/31/2019 and open vial date is 5 days ago.The meter memory was reviewed for previous test result history.(b)(6).Customer states that her doctor put her on two new insulin.Customer run a back to back test and got 50-point different on the same hand same drop of blood.I review with customer about proper back to back testing and customer states that she in unable trust the meter.Customer states that some of her results are high blood and erratic customer normal glucose range is 90-100mg/dl fasting, and she test 2 times day.Strips are being stored in the bathroom.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with tests results from all the results obtained.The expected fasting; blood glucose test result range is 90-100mg/dl, the customer did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bathroom.During the call on 11/12/2018,, a blood test was not performed by the customer.The test strip lot manufacturer's expiration date is 08/31/2019 and open vial date is 5 days ago.The meter memory was reviewed for previous test result history.Result 1:155mg/dl, date:11/12/2018, time:852am, fasting; result 2:133mg/dl, date:11/12/2018, time852am, fasting; result 3:296 mg/dl, date:11/11/2018, time:206am, fasting; result 4:346mg/dl, date:11/11/2018, time:205am, fasting; result 5:164mg/dl, date:11/11/2018, time: 856pm, fasting.Customer states that her doctor put her on two new insulin.Customer run a back to back test and got 50point different on the sam, e hand sam,e drop of blood.I review with customer about proper back to back testing and customer states that she in unable trust the meter.Customer states that some of her results are high blood and erratic customer normal glucose range is 90-100mg/dl, fasting; and she test 2 times day.Strips are being stored in the bathroom.
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4) returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-20-user's test strip had poor storage(bathroom).Test strip udi#: (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8135457
Report Number1000113657-2018-01218
Device Sequence Number1686086
Product Code NBW
UDI-Device Identifier00021292006051
UDI-Public(01)00021292006051
Combination Product (Y/N)N
Initial Reporter StateAR
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2802S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 11/12/2018
Supplement Date Received by Manufacturer11/12/2018
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-