Model Number 2430-0500
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Materials management; healthtrust analyst, member support.The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event or Problem Description
The customer reported the set leaked blood at the connector during patient use.There was no patient harm.
 
Additional Manufacturer Narrative
The customer¿s report of a leak at the connector was not confirmed; however, a leak was found at the top of the silicone pump segment.Functional testing found that the set leaked from the top of the silicone segment.Closer inspection under magnification observed that the leak is due to a small hole on the silicone segment.The cause of the leak is from a small hole in the silicone pump segment near the upper fitment.The root cause of the damage could not be definitively determined.
 
Event or Problem Description
The customer reported the set leaked blood at the connector during patient use.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key8135762
Report Number9616066-2018-02418
Device Sequence Number1721580
Product Code FPA
UDI-Device Identifier07613203011396
UDI-Public7613203011396
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K944320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2021
Device Model Number2430-0500
Device Catalogue Number2430-0500
Device Lot Number18067163
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date02/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-