| Model Number |
2430-0500 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Materials management; healthtrust analyst, member support.The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
The customer reported the set leaked blood at the connector during patient use.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a leak at the connector was not confirmed; however, a leak was found at the top of the silicone pump segment.Functional testing found that the set leaked from the top of the silicone segment.Closer inspection under magnification observed that the leak is due to a small hole on the silicone segment.The cause of the leak is from a small hole in the silicone pump segment near the upper fitment.The root cause of the damage could not be definitively determined.
|
| |
|
Event or Problem Description
|
|
The customer reported the set leaked blood at the connector during patient use.There was no patient harm.
|
| |
|