• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 12 DEFIBRILLATOR/MONITOR SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 12 DEFIBRILLATOR/MONITOR SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 99400-002505
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and could not duplicate the reported issue.Physio completed unrelated repairs and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
 
Event or Problem Description
The customer contacted physio-control to report that the device is unable to monitor paddles ecg.As a result, defibrillation therapy may not be available to a patient if needed.There was no patient use associated with the reported issue.
 
Additional Manufacturer Narrative
Physio further evaluated the reported issue and determined the report of the device being unable to monitor paddles ecg was recorded in error.The hospital biomed did not recall any issues with the displayed paddles lead ecg.
 
Event or Problem Description
The customer contacted physio-control to report that the device is unable to monitor paddles ecg.As a result, defibrillation therapy may not be available to a patient if needed.There was no patient use associated with the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 12 DEFIBRILLATOR/MONITOR SERIES
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key8135920
Report Number0003015876-2018-02065
Device Sequence Number16685916
Product Code MKJ
UDI-Device Identifier00885074269252
UDI-Public00885074269252
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K102972
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number99400-002505
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2018
Initial Date Received by Manufacturer 11/08/2018
Supplement Date Received by Manufacturer01/14/2019
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date01/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-