Model Number ZXT150
Medical Device Problem Codes Mechanical Problem (1384); Positioning Problem (3009)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided; best estimate is between (b)(6) and (b)(6) 2018.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxt150 22.0 diopter intraocular lens (iol) was implanted in the patient's right eye (od) on (b)(6) 2018, and later explanted on (b)(6) 2018 due to multiple adverse symptoms.Patient's visual acuity pre-operative was 20/30.Patient experienced halos and glare (dysphotopsia), blurry vision, and myopia after the procedure.The customer also indicated mechanical complication and positioning issue with the lens.Patient's visual acuity post-operative is 20/60.The explanted lens was replaced with another zxt150 of a lower diopter (20.5).No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: product testing could not be performed because the product was not returned; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search of complaints revealed no additional investigation requests have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8136163
Report Number9614546-2018-01137
Device Sequence Number1448382
Product Code POE
UDI-Device Identifier05050474605244
UDI-Public(01)05050474605244(17)230102
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/02/2023
Device Model NumberZXT150
Device Catalogue NumberZXT150U220
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/06/2018
Supplement Date Received by Manufacturer12/17/2018
10/25/2020
Initial Report FDA Received Date12/05/2018
Supplement Report FDA Received Date12/31/2018
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;