Model Number MMT-551NAP
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump received with intermittent button response due to flattened express bolus, esc, act, up and down button dome switch (creased).No unlocked j2/lcd keypad connector.Pump received with cracked case at the reservoir tube window corners.
 
Event or Problem Description
The customer reported via phone call that insulin pump had cosmetic damage.The blood glucose at the time of the incident was unknown.The customer was assisted with troubleshooting.The device will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8136429
Report Number2032227-2018-71308
Device Sequence Number262707
Product Code OZO
UDI-Device Identifier00643169736047
UDI-Public(01)00643169736047
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551NAP
Device Catalogue NumberMMT-551NAP
Device Lot NumberA6551NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2018
Initial Report FDA Received Date12/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
Patient Weight165
-
-