| Model Number |
MMT-723LNAS |
| Medical Device Problem Codes |
No Display/Image (1183); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a blank display.The patient¿s blood glucose level was 151 mg/dl at the time of the incident.Issue was not resolved by troubleshooting.Customer stated the display did not return.The customer was advised to discontinue use of the insulin pump and to revert to the backup plan.The device was not expected to be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin ump monitored for several days and no blank display noted.Insulin pump received with normal operating current.Insulin pump passed self test.Insulin pump passed off no power test.
|
| |
|
Search Alerts/Recalls
|