| Model Number |
Z9002 |
| Medical Device Problem Code |
Optical Discoloration (2999)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that patient had a blurred vision approximately after 4-month post-implantation of model z9002 tecnis monofocal iol (intraocular lens) in patient's left eye (os).Doctor reportedly noticed that lens had become dark.As a result, iol was explanted as planned.Initial post-surgical va 10/10.Patient complaining of blurred vision approximately 4 months post-surgical; however, the va was 10/10.The replacement lens was a non-jjsv iol.Patient is doing fine with perfect vision post-replacement.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation; returned to manufacturer on: 12/3/2018.Device evaluation: photos and details of the products used in cataract surgery for suspect tecnis cl z9002 lenses was provided.Research and development (r&d) performed an evaluation of similar issue reported and how this can affect the event.R&d chemistry wrote a technical report in which testing was done to the returned lens.As part of the results obtained from this report it was found that the lens when hydrated had a cloudy lens rating between 3 and 4.The haze level acceptance criteria is levels 1 to 4.Therefore, no product quality deficiency was observed.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no additional investigation requests were received for this production order.Labeling review: the labeling review was completed.The directions for use (dfu) adequately provide instructions, precautions, along with warnings for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|