Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018 and (b)(6) 2018.If explanted, give date: not applicable, as the lens remains implanted.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
It was reported that a zkb00 15.5 diopter lens was implanted in the patient's left eye (os) on (b)(6) 2018.The lens remains implanted, but an ar40 model lens with 1.0 diopter was implanted in the patient's left eye as a piggyback.There was no incision enlargement and no vitrectomy.Reportedly, the patient is doing okay.Intraocular lens (iol) master results indicated that the a constant used was correct (119.3 srkt).No additional information was provided.
|