Model Number ZCT600
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If explanted, give date: not applicable as the lens remains implanted.Device evaluation: the product testing could not be performed as the intraocular lens remains implanted.The customer's reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the zct600 intraocular lens (iol) was implanted into the patient's right eye at 006 degrees and was measured at the 1-month post-op visit at 180 degrees.There was no patient injury and the iol remains implanted.The outcome does not significantly interfere with the patient's activities of daily life and there is currently no plan to reposition the iol.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8136688
Report Number9614546-2018-01141
Device Sequence Number11240264
Product Code HQL
UDI-Device Identifier05050474604186
UDI-Public(01)05050474604186(17)201021
Combination Product (Y/N)Y
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional,study,use
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2020
Device Model NumberZCT600
Device Catalogue NumberZCT600U210
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/07/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Date Report Sent to FDA01/01/2005
-
-