• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-30-00
Medical Device Problem Codes Device Alarm System (1012); Pumping Stopped (1503)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event or Problem Description
Livanova (b)(4) received a report that a s5 system alarmed without warning massages and that a pump stopped during procedure.There was no report of patient injury.
 
Additional Manufacturer Narrative
Through further provided information livanova deutschland determined that the reported issue will now considered as not reportable.As a pump stop which was initiated on purpose during a demonstration/training and a following error which was not displayed on the central display is not likely to result in death or serious injury.
 
Event or Problem Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8137172
Report Number9611109-2018-01374
Device Sequence Number199153
Product Code DTQ
UDI-Device Identifier04033817900887
UDI-Public(01)04033817900887(11)170707
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48-30-00
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/08/2018
Supplement Date Received by Manufacturer04/05/2019
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date05/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-