| Model Number |
BTC93 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: visual analysis: visual inspection shows evidence of a crack at the base of the "y" connector.Performance analysis: pressure integrity testing was performed at 3 l/pm with 23 psi, (1189 mmhg) of back pressure for 10 minutes.During the pressure integrity testing, there was a small leak observed at the base of the "y" connector.Conclusion: the leakage reported only occurs in those assemblies that used cavity 1.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use the suction and anticoagulant line leaked when the customer hooked it up to saline.The product was replaced.There was no patient involved so no adverse effect occurred.
|
| |
|