MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Codes High Test Results (2457); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f).Internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-18- user has high glucose value.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for hi display.The expected fasting blood glucose test result range is undisclosed.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the living room.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of hi display using true metrix meter.The test strip lot manufacturer's expiration date is 02/29/2020 and open vial date is yesterday.The meter memory was reviewed for previous test result history.Hi display 11/ 1113 820a fasting.560mg/dl 11/12 1046a fasting.Customer only has two readings.Customer called regarding e-2 error message.Ccr rep educated customer and attempted to trouble shoot.While performing a blood test customer got hi as a result.Customer stated that he does not have any symptoms.Customer is storing strips properly.Customer was release from the hospital recently due to kidney transplant, customer is also on steroid medication for the surgery.Customer tested while in the hospital and blood sugar was in the 500'smg/dl.Customer received our meter yesterday and tested getting a reading of 560mg/dl fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8137491
Report Number1000113657-2018-01221
Device Sequence Number1579513
Product Code NBW
UDI-Device Identifier00021292006051
UDI-Public(01)00021292006051
Combination Product (Y/N)N
Initial Reporter StateCT
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2920S
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2018
Initial Report FDA Received Date12/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-