• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR Back to Search Results
Model Number E SERIES
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while monitoring a patient (age & gender unknown) during transport in an ambulance, the device inappropriately shut down.Complainant indicated that the clinician plugged the device to the inverter in the ambulance to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation evaluated the device and the reported malfunction was not observed with the device.The device was put through extensive testing which included bench handling and stress testing without duplicating the malfunction.The device's high voltage battery pins were replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E SERIES DEFIBRILLATOR
Common Device Name
E SERIES
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8137513
Report Number1220908-2018-03509
Device Sequence Number13502520
Product Code MKJ
UDI-Device Identifier00847946015671
UDI-Public00847946015671
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K042007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberE SERIES
Device Catalogue NumberE SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-