| Model Number |
METRIX AIR |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Headache (1880)
|
| Date of Event |
11/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-62-user had poor technique test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product notification letter to customer to contact customer care.No address on file.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-3 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of a headache.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored according to specification in the bedroom.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of 152mg/dl non-fasting using true metrix air meter.The test strip lot manufacturer's expiration date is 05/17/2020 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = t) internal report # (b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Mlc-30 user applied blood to strip before inserting strip into meter test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to sendvproduct notification letter to customer to contact customer care.No address on file.Correction: correct mlurc on section h10 after reviewing complaint.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-3 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of a headache.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored according to specification in the bedroom.During the call on 11/13/2018, a blood test was performed by the customer and produced test results of 152mg/dl non-fasting using true metrix air meter.The test strip lot manufacturer's expiration date is 05/17/2020 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|