| Model Number |
G22161 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
This file is being reported based on the fda mdr malfunction precedence "stent kinked/ bent" stent kinked."as per complaint form": the assistant will straight the pigtail stent with the straighter, but the stent is kinking.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining (b)(4).Importer site establishment registration number: (b)(4).1.Device evaluation: 1 x zso-7-4 of lot number c1381597 was returned to cirl for an evaluation.It was in a used state and it was returned in it¿s opened original packaging.2.Lab evaluation: 1 x zso-7-4 of lot number c1381597 was returned to cirl for an evaluation on 22nd of november 2018.In summary the following results were observed in the lab evaluation.Kinks were observed at the 2nd, 4th & 5th porthole on the tapered end of the stent.A kink was observed at the 4th porthole on the non-tapered end of the stent.Straighter on stent- flared side is facing the non-tapered end.Complaint is confirmed as the failure was verified in the laboratory.3.Root cause (possible): a definite root cause cannot be determined, as the circumstances of use cannot be replicated in the laboratory.A possible cause of this complaint is that the kink occurred while being straightened while being passed over wire guide.A precaution included in instructions for use, ifu0045-6 states "care must be exercised when straightening the pigtail curls* in order to avoid kinking or breaking the stent." 4.Documents review: a review of the manufacturing records for the device of lot # c1381597 did not reveal any discrepancy related to the complaint issue.Prior to distribution all zimmon bilary stent set zso-7-4 are subjected to visual and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place in cirl.The instructions for use, ifu0045-6 which accompanies this device instructs the user to ¿visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ also the user is instructed by the instructions for use, ifu0045-6 ¿care must be exercised when straightening the pigtail curls in order to avoid kinking or breaking the stent.¿ the review of relevant information provided, there is no evidence to suggest that the user did not follow the instructions for use.5.Summary: complaint is confirmed as the failure was verified in the laboratory.The patient didn¿t experience any adverse effects as result of the device failure and based on the information provided.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
This file is being reported based on the fda mdr malfunction precedence "stent kinked/ bent" stent kinked."as per complaint form": the assistant will straight the pigtail stent with the straighter, but the stent is kinking.
|
| |
|
Search Alerts/Recalls
|