MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HD 22 + 8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HD 22 + 8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71302208
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 11/15/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an open reduction was performed due to dislocation.No delay or additional injuries reported.
 
Additional Manufacturer Narrative
The associated r3 constrained liner and cocr femoral head were not returned for evaluation.Therefore products evaluation and dimensional analysis could not be performed.However, device details were provided.Thus, a review of device history record was performed.The review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.Our investigation indicated that the x-ray provided confirms the dislocation.However, without the request clinical information, fall history, primary operative report, open reduction procedure report, or the explant the root cause of the dislocation cannot be concluded.The further impact to the patient beyond the revision cannot be determined.No further clinical/medical assessment is warranted at this time.Without the return of the actual products involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the devices or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HD 22 + 8
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8137588
Report Number1020279-2018-02698
Device Sequence Number18880651
Product Code LPH
UDI-Device Identifier03596010194053
UDI-Public03596010194053
Combination Product (Y/N)N
PMA/510(K) Number
K963509
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/20/2019
Device Catalogue Number71302208
Device Lot Number09BM16931
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date05/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-