Catalog Number 6260-5-228
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Injury (2348); Reaction (2414); Ambulation or Postural Difficulties (2544)
Date of Event 11/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding altr involving a metal head was reported.The event was not confirmed.Method & results: device evaluation and results:material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the event could not be confirmed and exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, histopathology report, progress notes and x-rays as well as device return and evaluation are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
It was reported that patient's right hip was revised due to metallosis after patient complaint of pain and difficulty ambulating.A 28 +4 head and 28 10° e liner were revised to a 32 0° liner and ceramic head with sleeve.A restoration ps stem and trident psl cluster shell were not revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
28MM +4 V40 TAPER VIT HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8137848
Report Number0002249697-2018-03928
Device Sequence Number16621850
Product Code JDI
UDI-Device Identifier07613327012897
UDI-Public07613327012897
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2013
Device Catalogue Number6260-5-228
Device Lot Number25877801
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date12/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight74
-
-