| Model Number |
27888 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Therapeutic Effects, Unexpected (2099)
|
| Date of Event |
11/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the patient went to have a revision due to the product ¿not working.¿ during the revision, the physician found that the valve was leaking from the delta chamber.It was stated there were no external/environmental/patient factors identified that may have led to the issue.The valve was replaced and new valve seemed to be working fine.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for valve flux, siphon, pressure-flow and preimplantation.It did not meet the requirements for leak and reflux testing due to tears on the delta chamber casing and distal occluder.The instructions for use cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity¿¿ there was debris on the interior and exterior of the valve.The instructions for use cautions, ¿care must be taken to ensure that particulate contaminants are not introduced into shunt components during preimplantation testing or handling.Introduction of contaminants could result in improper performance of the shunt system.Particulate matter that enters the shunt system may result in shunt occlusion or may also hold pressure/flow controlling mechanisms open, resulting in overdrainage.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|