• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10601
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that shaft break occurred.A 4.00 x 12mm synergy ii drug-eluting stent was advanced for treatment; however, it was noted that the main body was fractured more than 15cm from the hub, resulting in two separate pieces of the stent delivery system (sds).It was noted under fluoroscopy that the distal end of the stent never exited the guiding catheter and so the entire fractured sds was removed together with the guide catheter.The procedure was completed with another of the same device.No patient harm nor complications were reported.
 
Event or Problem Description
It was reported that shaft break occurred.A 4.00 x 12mm synergy ii drug-eluting stent was advanced for treatment; however, it was noted that the main body was fractured more than 15cm from the hub, resulting in two separate pieces of the stent delivery system (sds).It was noted under fluoroscopy that the distal end of the stent never exited the guiding catheter and so the entire fractured sds was removed together with the guide catheter.The procedure was completed with another of the same device.No patient harm nor complications were reported.
 
Additional Manufacturer Narrative
Device is a combination product.Device evaluated by mfr: synergy ii us mr 4.00 x 12mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent od(outer diameter) was measured and was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no damage.A visual and tactile examination of the hypotube found a break 725mm distal to the distal end of the strain relief.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8138528
Report Number2134265-2018-63472
Device Sequence Number262704
Product Code NIQ
UDI-Device Identifier08714729840077
UDI-Public08714729840077
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2020
Device Model Number10601
Device Catalogue Number10601
Device Lot Number0022411480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer12/20/2018
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date01/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-