SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH DEVICE & POWER SUP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801280
Device Problem Failure to Charge (1085)
Patient Problem Perforation of Esophagus (2399)
Event Date 11/09/2018
Event Type  malfunction  
Event Description
It was reported that during use the console does not charge at all.No patient harm was reported.
 
Manufacturer Narrative
Our investigation into the complaint has now concluded.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.The device was sent to our technical centre so a thorough technical analysis could be carried out to help to identify if there were any problems and/or what could have caused the reason for the complaint.During functional evaluation of the device the complaint issue of failure to charge was confirmed as the battery would not fully charge.The cause of this was identified as a faulty battery.Following the complaint assessment suitable repairs were conducted and the device was tested and confirmed working within expected parameters.Smith and nephew are continually investigating ways to develop and improve our products and we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH DEVICE & POWER SUP
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key8139036
MDR Text Key129582269
Report Number8043484-2018-00427
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801280
Date Manufacturer Received11/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-