| Model Number |
0230-14-1.7 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.Upon completion of the sample evaluation and investigation; a follow-up report will be filed.All information reasonably known as of 06-dec-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a jejunostomy tube broke off in the patient.Additional information received on (b)(6) 2018 stated that the tube was placed on (b)(6) 2018.The facility staff believed that the patient cut the tube.However, there could have also been a clog in the tube which resulted in it bursting from the pressure of adding fluid.The distal portion of the tube was removed via endoscopy.A new tube was placed with no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Avanos medical received one used sample.No product packaging was received.The sample was evaluated and the reported failure was confirmed.The cause for this incident seems to be outside of the scope of the manufacturing process.The root cause of the reported issue has not been conclusively determined.The device history record for the reported lot number was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.All information reasonably known as of 04-feb-2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|