WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/10 HOLE/230MM/RIGHT; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.124.410
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is an unknown date in (b)(6) 2018.Complainant part is not expected to be returned for manufacturer review/investigation.Date of concomitant therapy is the same as date of implantation, an unknown date in (b)(6) 2018.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, patient underwent an unknown surgery.The variable angle locking compression plate (va-lcp) curved condylar plate was attached to an aiming arm.While the surgeon was attempting to attach percutaneous trocar to the jig, the aiming arm fell off the plate.This occurred a few times during the case until the jig was no longer able to grab the plate; presumably the hole in the plate was stripped out.Upon pulling the plate, the head of a cortex screw self-tapping broke off.They were able to retrieve the screw fragment in the medullary canal with some difficulty.A new, longer plate was placed and no other problems occurred.Procedure outcome is unknown.There was no patient consequence.Concomitant devices: trocar instrument (part: unknown, lot: unknown, quantity: 1), insertion handle (part: unknown, lot: unknown, quantity: 1), aiming arm (part: unknown, lot: unknown, quantity: 1).This report is for a 4.5mm va-lcp curved condylar plate.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: jds, hwc.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Occupation: reporter is a synthes employee.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/10 HOLE/230MM/RIGHT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8139263
Report Number2939274-2018-55311
Device Sequence Number19272363
Product Code HRS
UDI-Device Identifier10886982042811
UDI-Public(01)10886982042811
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.124.410
Device Catalogue Number02.124.410
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/06/2018
Supplement Date Received by Manufacturer12/06/2018
12/06/2018
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date12/06/2018
12/06/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age15 YR
Patient Weight60
-
-