WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/10 HOLE/230MM/RIGHT; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
02.124.410 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is an unknown date in (b)(6) 2018.Complainant part is not expected to be returned for manufacturer review/investigation.Date of concomitant therapy is the same as date of implantation, an unknown date in (b)(6) 2018.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, patient underwent an unknown surgery.The variable angle locking compression plate (va-lcp) curved condylar plate was attached to an aiming arm.While the surgeon was attempting to attach percutaneous trocar to the jig, the aiming arm fell off the plate.This occurred a few times during the case until the jig was no longer able to grab the plate; presumably the hole in the plate was stripped out.Upon pulling the plate, the head of a cortex screw self-tapping broke off.They were able to retrieve the screw fragment in the medullary canal with some difficulty.A new, longer plate was placed and no other problems occurred.Procedure outcome is unknown.There was no patient consequence.Concomitant devices: trocar instrument (part: unknown, lot: unknown, quantity: 1), insertion handle (part: unknown, lot: unknown, quantity: 1), aiming arm (part: unknown, lot: unknown, quantity: 1).This report is for a 4.5mm va-lcp curved condylar plate.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: jds, hwc.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Occupation: reporter is a synthes employee.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|