• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TISSUE SCIENCE LABORATORIES MESH TSL - PERMACOL¿; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TISSUE SCIENCE LABORATORIES MESH TSL - PERMACOL¿; MESH, SURGICAL Back to Search Results
Model Number P151828
Medical Device Problem Codes Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Adhesion(s) (1695); Bacterial Infection (1735); Purulent Discharge (1812); Fistula (1862); Unspecified Infection (1930); Perforation (2001); Scar Tissue (2060); Scarring (2061); Discharge (2225); Injury (2348); Impaired Healing (2378); Obstruction/Occlusion (2422); Fluid Discharge (2686); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced chronic wound infection with infected hernia mesh.Post-operative patient treatment included revision surgery/mesh removal.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after underlay implant, the patient experienced chronic wound infection with infected hernia mesh, fistula, chronic draining sinuses, adhesions, serosal tears, mesh falling apart and heavily degraded, scar, free pus, ileus, strep infection, and open wound.Post-operative patient treatment included revision surgery/mesh removal, debridement of abdominal wall incision, lysis of adhesions, enterorrhaphy, antibiotics, and wound vac.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced chronic wound infection with infected hernia mesh, fistula, and open wound.Post-operative patient treatment included revision surgery/mesh removal, debridement of abdominal wall incision, and wound vac.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.Additional information: a2, a4 (weight in lbs), b5, b7, e3, e4 (email), g4, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after underlay implant, the patient experienced chronic wound infection with infected hernia mesh, fistula, chronic draining sinuses, adhesions, serosal tears, mesh falling apart and heavily degraded, scar, free pus, ileus, strep infection, open wound, pain, perforation, bad odor from wound.Post-operative patient treatment included revision surgery/mesh removal, debridement of abdominal wall incision, lysis of adhesions, enterorrhaphy, antibiotics, wound vac.Relevant tests/laboratory data (b)(6) 2014: cultures positive for strep.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESH TSL - PERMACOL¿
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB  GU11 1EJ
Manufacturer (Section G)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB   GU11 1EJ
Manufacturer Contact
justin ellis
60 middletown ave
8200 coral sea st ne
mounds view,mn, CT 55112
7635265677
MDR Report Key8139527
Report Number9617613-2018-00117
Device Sequence Number13889600
Product Code FTM
UDI-Device Identifier10884523000191
UDI-Public10884523000191
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120605
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2014
Device Model NumberP151828
Device Catalogue NumberP151828
Device Lot Number11B0808
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/14/2018
Supplement Date Received by Manufacturer06/26/2019
06/26/2019
08/24/2022
09/07/2022
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date01/14/2020
02/04/2020
08/29/2022
09/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
Patient Weight101 KG
-
-