• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HIGH OFFSET NEUTRAL MODULAR NECK; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HIGH OFFSET NEUTRAL MODULAR NECK; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71352112
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 02/13/2014
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient underwent an aspiration on the left hip due to an infection.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and although an aspiration was performed to rule out an infection prior to the left hip revision, it was unsuccessful and ¿prematurely terminated¿ per patient request.The pathology results of the left hip revision intra-op specimens revealed ¿fibroligamentous tissues¿ which are consistent with an alval/pseudotumor.The patient impact beyond the reported discomfort and anxiety during the attempted aspiration cannot be determined.No further medical assessment can be rendered at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH OFFSET NEUTRAL MODULAR NECK
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8139688
Report Number1020279-2018-02705
Device Sequence Number1593717
Product Code LPH
UDI-Device Identifier03596010619549
UDI-Public03596010619549
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K072417
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2014
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/01/2022
Device Catalogue Number71352112
Device Lot Number12GTM0010D
Initial Date Received by Manufacturer 11/07/2018
Supplement Date Received by Manufacturer11/07/2018
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date04/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age64 YR
Patient Weight82
-
-