| Catalog Number |
48132601 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Fall (1848); Pain (1994); Injury (2348); Ambulation or Postural Difficulties (2544)
|
| Date of Event |
05/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the rod broke in half and screws came out post-operatively.Patient reportedly had fall in (b)(6) of 2017 in which she tripped over a chair and fell on their knees.The patient was in pain, couldn't walk, trouble breathing and couldn't lift arm.The patient underwent revision surgery on (b)(6) 2018 to replace hardware.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: surgical notes review, x-ray image review, nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed upon review of surgical notes and x-ray image.Products were not returned for evaluation and no lot# was provided therefore manufacturing records review and complaint history review could not be performed.Patient was reported to have had a fall in thanksgiving of 2017 in which she tripped over a chair and fell on knees.Conclusion: the root cause of the reported event could not be determined conclusively but it is likely that the patient fall (a few months ago) may have contributed to the event - perhaps a crack in the rod at the time of fall which may have lead to fracture due to fatigue over time.
|
| |
|
Event or Problem Description
|
|
It was reported that the rod broke in half and screws came out post-operatively.Patient reportedly had fall in thanksgiving of 2017 in which she tripped over a chair and fell on their knees.The patient was in pain, couldn't walk, trouble breathing and couldn't lift arm.The patient underwent revision surgery on (b)(6) 2018 to replace hardware.
|
| |
|
Search Alerts/Recalls
|