| Catalog Number |
1005203 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an unspecified artery in the right leg below the knee that did not have any tortuosity or calcification.A 300 cm spartacore guide wire was used; however, the tip separated inside the anatomy during removal with a support catheter.Therefore, a snare device was used to remove the component, and another unspecified spartacore guide wire was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|